The TinyTag Carbon Dioxide Data Logger monitors and logs carbon dioxide levels using an infrared sensor. It can be used in experimental situations or simply to monitor air quality within buildings to ensure that ventilation is effective. The logger possesses a self-calibrating NDIR sensor consisting of an infrared source. A second infrared source is powered up occasionally to calibrate the primary sensor and this patented calibration technique ensures excellent long-term accuracy.
The Carbon Dioxide Logger is housed in a discreet case and is powered by the mains using the power adaptor provided. In the event of a power failure a red LED informs the user that there is a problem. Two models of the Tinytag Carbon Dioxide Data Logger are available. The standard version, TGE-0010, has a measurement range of 0-2000ppm and the wide-range version, TGE-0011, is capable of measuring 0-5000ppm.
To download data from this logger a USB cable and Explorer Software will also be required.
Key features:
* Total reading capacity: 32,000 readings
* Memory type: Non-volatile
* Delayed start: Relative/absolute (up to 45 days)
* Stop options: When full / After n readings / Never (overwrite oldest data)
* Reading types: Actual / min / max
* Logging interval: 1 min to 10 days
* Alarms: 1 fully programmable (a second alarm can be programmed at the expense of the “no power” warning)
* One year warranty
Reading specification:
* Sensor type: NDIR Infra-red CO2 sensor
* Response time t63: < 195s
* Temperature dependence: 2ppm CO2/°C
* Long term stability: 20ppm/year (typical)
* Reading resolution: 0.1ppm
* TGE-0010 accuracy: < ± 50ppm +2% of measured value
* TGE-0011 accuracy: < ± 50ppm +3% of measured value
The accuracy figures quoted above are at 25°C (77°F) and 1013mbar
A warm-up time of 5 minutes is required to achieve values specified above
Physical specification:
* Working temperature range: -20°C to +60°C (-4°F to +140°F)
* Case dimensions: Width 85mm, Height 100mm, Depth 26mm
* Weight: 100g
Power specification:
* The unit is supplied with a mains adapter that comes with UK, EU and US AC connectors
* Back-up battery type: CR2325
* Before replacing the battery the data logger must be stopped
* Data stored on the logger will be retained after a battery is replaced
* The unit also has a +18V DC screw terminal connection on its circuit board that can be used to power the unit from a separate supply if required